Continuity · Particle Control
ISO 14644 Particle Control
A particle counter in a room gives you a number. ISO 14644 turns that number into documentation: the right class, the right sampling locations, the right occupancy state and a plan that detects changes in time. We classify, find the sources and build the monitoring that keeps particle levels under control in operation.
- 9
- Cleanliness classes in ISO 14644-1, from ISO Class 1 as the cleanest to ISO Class 9
- 24
- Sampling locations is the minimum for a room of up to 352 m² according to Table A.1
- 3
- Occupancy states: as-built, at-rest and operational
- 2 litres
- Minimum sample volume per location, sampled for at least 1 minute
The standard
What is ISO 14644?
ISO 14644 is the international series of standards for cleanrooms and associated controlled environments. Part 1 classifies air cleanliness by the concentration of airborne particles at specified particle sizes.
The series covers the whole life of a controlled area: from design and start-up through classification and testing to monitoring and daily operation.
- BuildISO 14644-4: design, construction and start-up
- ClassifyISO 14644-1: cleanliness class by particles
- TestISO 14644-3: test methods for performance
- MonitorISO 14644-2: evidence of performance over time
- OperateISO 14644-5: requirements for daily operation
| Part | Subject | Used for |
|---|---|---|
| ISO 14644-1:2015 | Classification of air cleanliness by particle concentration | The class, sampling locations and sample volume |
| ISO 14644-2:2015 | Monitoring to provide evidence of cleanroom performance | The monitoring plan in operation |
| ISO 14644-3:2019 | Test methods | Investigating airflow and filters, for example |
| ISO 14644-4:2022 | Design, construction and start-up | New and rebuilt areas |
| ISO 14644-5:2025 | Operations | Behaviour, cleaning and materials in daily operation |
| ISO 14644-8:2022 | Air cleanliness by chemical concentration | Chemical air contamination |
| ISO 14644-9:2022 | Surface cleanliness by particle concentration | Particles on surfaces |
| ISO 21501-4 | Calibration of airborne particle counters | The particle counter used for the measurement |
Cleanliness classes
Nine classes, each ten times cleaner
The class states the maximum permitted number of particles per m³ of air at the particle sizes included in the measurement. The limits are cumulative, so the figure at 0.5 µm includes all particles of 0.5 µm and larger.
From one class to the next, the limit changes by a factor of ten. ISO Class 8 therefore permits ten times as many particles as ISO Class 7.
Particles ≥ 0.5 µm per m³, logarithmic scale
| Class | ≥ 0.5 µm | ≥ 1 µm | ≥ 5 µm |
|---|---|---|---|
| ISO Class 5 | 3,520 | 832 | Not specified |
| ISO Class 6 | 35,200 | 8,320 | 293 |
| ISO Class 7 | 352,000 | 83,200 | 2,930 |
| ISO Class 8 | 3,520,000 | 832,000 | 29,300 |
| ISO Class 9 | 35,200,000 | 8,320,000 | 293,000 |
The limits are calculated as Cn = 10N × (0.1/D)2.08, where N is the class and D the particle size in µm. Intermediate classes can be specified in steps of 0.1. ISO Class 9 applies only in the operational occupancy state.
In data centres, ISO Class 8 is the usual target after technical cleaning. Read more under data centre and server room cleaning.
Classification
The test basis is set before the measurement
Two particle measurements can only be compared if the sampling locations, the instrument, the occupancy state and the activity are known. That is why the test basis is agreed before the particle counter is switched on.
The number of sampling locations follows from the area of the room. Their placement is set according to the layout, and critical spots can require extra locations.
- State 1As-built
The installation is complete, with all services connected and running. No production equipment, no materials and no people.
- State 2At-rest
The equipment is installed and running as agreed, but there are no people in the room.
- State 3Operational
The room is in operation with the agreed staffing and activity. The hardest state to pass.
- Classification location (24)
- Diagnostic location: A airlock, B packing station, C assembly station by a door
| Area up to | Locations |
|---|---|
| 10 m² | 5 |
| 24 m² | 6 |
| 52 m² | 10 |
| 104 m² | 16 |
| 148 m² | 19 |
| 232 m² | 22 |
| 352 m² | 24 |
| 636 m² | 26 |
| 1,000 m² | 27 |

| Record | Why |
|---|---|
| Date and time | The measurement can be compared with operation and cleaning |
| Occupancy state | As-built, at-rest and operational give different levels |
| Number of people | People and their movements add particles |
| Activities and door openings | Fluctuations can be linked to what was going on |
| Cleaning and deviations | Explains single measurements that stand out |
| Sample volume | At least 2 litres and 1 minute per location, more for cleaner classes |
The case
From inconsistent measurements to a pattern
An electronics manufacturer saw a growing number of faults during assembly of sensitive components. The area had filtered air, access rules and fixed cleaning routines, but nobody could say whether airborne particles were a contributing cause.
The aim was to document the area to ISO Class 7 and keep particle control in place during normal production.
About the caseThe case is an illustrative, anonymised process. Company, measurements and results are constructed examples, not documented customer data. It shows how a project can be organised according to the principles of ISO 14644.
- 01Fixed methodA consistent classification method
- 02LocationsRepresentative and critical spots
- 03CausesConditions that explain high readings
- 04ImprovementsChanges and repeat measurements
- 05MonitoringA plan that detects changes in time
| Item | Choice in the case | Why |
|---|---|---|
| Class | ISO Class 7 | The requirement for assembly of sensitive components |
| Particle sizes | 0.5 µm and 5.0 µm | Fine particles and the larger ones that typically come from activity |
| Sampling locations | 24 distributed locations and 3 diagnostic locations | Table A.1 for about 300 m² plus airlock, packing station and an assembly station |
| Occupancy states | At-rest and operational | The difference shows what operation itself adds |
Baseline: the fluctuations followed specific activities
Most locations were stable at 0.5 µm, but in some places the concentration rose clearly during operation: at the packing station when packaging was opened, at the airlock when materials were moved in, at an assembly station near a frequently used door and shortly after dry cleaning of a work surface. The figures are fictitious and not a classification report.
| Area | 0.5 µm | 5.0 µm | Observation |
|---|---|---|---|
| Assembly stations away from doors | 125,000–210,000 | 300–1,500 | Mostly stable values |
| Packing station | 180,000–310,000 | 1,200–3,600 | Short peaks when packaging is opened |
| Airlock and material route | 160,000–280,000 | 1,000–3,300 | Fluctuations with transport and door openings |
| Assembly station near door | 170,000–290,000 | 1,100–3,400 | Raised level with simultaneous traffic |
The 5.0 µm limit for ISO Class 7. Three of the four areas exceeded it at their peaks. The pattern did not point to failing filtration of the whole room, but to material flow, work methods and local air movement.
- Packaging and materials
Packaging was opened close to the product, and some materials shed visible fibres.
- Traffic through the airlock
The door was sometimes held open while pallets and trolleys were moved through.
- Cleaning method
Dry cloths and quick movements can lift particles into the air instead of removing them.
- Air distribution
Air movement and ventilation were investigated as explanations for local fluctuations, using test methods from ISO 14644-3.
Improvements
Measures that can be measured one by one
The improvements were chosen because they could be linked directly to the observed fluctuations. That way the effect can be measured, instead of introducing general measures and hoping for the best.
Afterwards, the measurements were repeated with the same locations, the same staffing and comparable activity.

| Area | Measures |
|---|---|
| Material handling | Packaging opened in a separate zone away from assembly, waste removed continuously in closed containers, and materials wiped to a fixed procedure before entry |
| Airlock and traffic | Consolidated deliveries, separate flows for people and materials, never both airlock doors open at once, and fixed cleaning of trolleys |
| Cleaning and behaviour | New cleaning method, instruction in work movements, cleaning outside sensitive activities and recording of deviations |
| Technical review | Air distribution, filters and air supply assessed, and critical locations compared with the workstations |
Particles ≥ 5.0 µm per m³ during operation, before and after
- Assembly stations away from doors
- Packing station
- Airlock and material route
- Assembly station near door
- Before improvements
- After improvements
- Limit for ISO Class 7
The results were assessed across all locations and operating conditions, so that one good measurement did not hide a recurring problem with a specific activity.
Monitoring
From classification to continuous control
A classification documents the conditions during one measurement. It does not show that the area still meets the class when processes, staffing or equipment change.
ISO 14644-2 describes the minimum requirements for a monitoring plan that provides evidence of cleanroom performance over time.
- MeasureFixed locations at a fixed frequency
- RecordTogether with occupancy state and staffing
- AssessAgainst limits for follow-up
- InvestigateWhen a limit is exceeded
- AdjustAnd measure again after changes
| Element | Content |
|---|---|
| Fixed locations | At assembly areas and the packing station |
| Frequency | Set according to risk and process changes |
| Recording | Measurements together with occupancy state and staffing |
| Limits | Defined limits for follow-up and investigation |
| Extra measurement | After changes to ventilation, cleaning, production flow or layout |
| Periodic review | Are the locations and frequency still relevant? |
Scope
What a particle measurement does not show
Particle counting documents the concentration of airborne particles at the sizes included in the measurement. It says nothing about the particles' material, chemical content or biological nature.
An ISO class is not an overall approval of the process, a guarantee of product quality or documentation of GMP compliance either.
- ShowsAirborne particles
Number per m³ at the measured sizes
ISO 14644-1 - Needs other methodsChemical air contamination
Gases and molecular contamination
ISO 14644-8 - Needs other methodsParticles on surfaces
Cleanliness of work surfaces and equipment
ISO 14644-9 - Needs other methodsMicroorganisms
Viable particles and microbiological conditions
Microbiological methods
Our service
How we work with particle control
We organise particle control according to ISO 14644, from the test basis to a monitoring plan your operation can live with. The measurements and reports are uploaded to your My page.
If the measurements show that the area needs cleaning, we carry out the technical cleaning and measure again afterwards.
- 01Test basisClass, particle sizes, occupancy state and locations
- 02BaselineClassification and diagnostic locations
- 03SourcesReview with operations and quality
- 04ImprovementsMeasures and technical cleaning
- 05RepeatSame locations and same set-up
- 06MonitoringA plan to ISO 14644-2

| Deliverable | Content |
|---|---|
| Classification report | Class, occupancy state, locations, sample volume and result per location |
| Source assessment | Fluctuations linked to activities, with prioritised measures |
| Repeat measurement | Before and after at the same locations |
| Monitoring plan | Locations, frequency, limits and when to measure extra |
| My page | All reports in one place, see Compliance & documentation |
In data centres and server rooms we combine the particle measurement with technical data centre cleaning and server room cleaning.
Questions
Frequently asked questions
Answers to what we are asked most often about ISO 14644 and particle measurement.
What is ISO 14644-1?
ISO 14644-1 is the part of the ISO 14644 series that classifies air cleanliness by the concentration of airborne particles. It defines nine classes from ISO Class 1 to ISO Class 9 and describes how sampling locations and sample volume are determined. The current edition is from 2015.
How many sampling locations are needed?
The minimum number follows from the room's area according to Table A.1 in ISO 14644-1. A room of up to 352 m² requires 24 locations, and a room of up to 1,000 m² requires 27. Critical spots can require extra locations in addition to the distributed ones.
Which ISO class should a data centre have?
In data centres, ISO Class 8 is the usual target after technical cleaning and before handover. The requirement depends on the equipment supplier's and the owner's requirements, so the class is agreed as part of the test basis.
What is the difference between classification and monitoring?
Classification to ISO 14644-1 documents cleanliness during one measurement. Monitoring to ISO 14644-2 provides ongoing evidence that the area continues to meet the class as operation changes.
Can a particle measurement show whether the air is free of bacteria?
No. A particle counter counts particles at specific sizes, but does not distinguish between viable and non-viable particles. Microbiological conditions require other methods.
How long must each location be sampled?
The sample volume must be large enough to statistically expect at least 20 particles at the class limit for the largest particle size, and at least 2 litres for at least 1 minute. The cleaner the class, the longer the sample.
Engineering continuity
Do you know where the particles come from?
We classify your area to ISO 14644, find the sources and build a monitoring plan that detects changes in time.